Beam Dynamics Simuation in the Injector for the Proton-Radiation Therapy Complex «Luch-Proton»
https://doi.org/10.25205/2541-9447-2024-19-4-20-30
Abstract
The “Luch-Proton” project of the proton beam therapy center is being created at the National Research Center «Kurchatov Institute». The pulsed proton accelerator RFQ with energy of 5 MeV, operating frequency of 162.5 MHz and current of 30 mA is being developed at the National Research Center “Kurchatov Institute” – KKTEF as an injector of the proton synchrotron. The main aspects of RFQ beam dynamics are described choosing RFQ parameters, optimizing the transport and transmission of the accelerating channel, minimizing the output beam emittance. The results of beam dynamics simulation in RFQ are presented.
About the Authors
G. N. KropachevRussian Federation
Gennady N. Kropachev, Ph.D. of Physico-mathematical Sciences, Leading Researcher
Moscow; Moscow region; Dubna
T. V. Kulevoy
Russian Federation
Timur V. Kulevoy, Doctor of Engineering Sciences, Deputy Head
KCTEP for applied scientific research and experimental installations
Moscow
A. L. Sitnikov
Russian Federation
Alexey L. Sitnikov, Ph.D. of Physico-mathematical Sciences, Senior Researcher
Moscow
E. R. Khabibullina
Russian Federation
Ekaterina R. Khabibullina, Junior Researcher
Moscow
References
1. Kapchinsky I. M. Theory of linear resonant accelerators. Мoscow, Energoizdat publ., 1982, pp. 130–144.
2. Kapchinsky I. M. About approximations of Kilpatrick criterion. PTE, 1986, № 1, pp. 33–35.
3. Kapchinsky I. M. Particle dynamics in linear resonance accelerators, M., 1966, pр. 247.
4. Wadlinger E. A. Scaling Laws for RFQ Design Procedures. Proc. PAC, 1985, Vancouver, British Columbia, May 13–16, 1985, pp. 2596–2598.
5. Kropachev G., Kulevoy T., Sitnikov A. The Proton Linac for Compact Neutron Source Daria, Journal of Surface Investigation: X-ray, Synchrotron and Neutron Techniques, 2019, vol. 13, no. 6, pp. 1126–1131. DOI: 10.1134/S1027451019060399
6. Kulevoy T. et al. Compact multipurpose facility – BELA. In Proc. 29 the Linear Accelerator Conf. (Beijing, China, Sep. 2018), p. 349. doi: 10.18429/JACoW-LINAC2018-TUPO012
7. Balabin A. I., Lipkin I. M., Ugarov S. B. On the choice of the shape of electrodes in a linear accelerator with spatially uniform quadrupole focusing. VANT. Ser.: Technique of physical experiment, Moscow, 1988, iss. 1(36), pp. 20–22.
8. Roshal A. S. Modeling of charged beams. Moscow, Atomizdat publ., 1979.
9. Uriot D., Pichoff N. Status of TraceWin Code, Proceedings of IPAC2015. Richmond, VA, USA, pp. 92–94.
Review
For citations:
Kropachev G.N., Kulevoy T.V., Sitnikov A.L., Khabibullina E.R. Beam Dynamics Simuation in the Injector for the Proton-Radiation Therapy Complex «Luch-Proton». SIBERIAN JOURNAL OF PHYSICS. 2024;19(4):20-30. (In Russ.) https://doi.org/10.25205/2541-9447-2024-19-4-20-30