Preview

SIBERIAN JOURNAL OF PHYSICS

Advanced search

Measurements of the Neutron Field Parameters with Monitors based on the Gas-Filled Proportional Counters

https://doi.org/10.25205/2541-9447-2024-19-2-41-49

Abstract

Neutron monitors with the gas filled proportional counters inside and first data were presented at RuPAC-2018 and RuPAC-2021 conferences. It is planned to use such devices to monitor the stability of irradiation conditions during radiation therapy sessions. We present here the calibration procedure for these monitors at AmBe neutron source, and the new data. These monitors have been used to measure the neutron fluence behind the shielding of the 450 MeV/nucleon carbon beam based experimental facility “Shared use Center – radiobiological stand with the carbon beam of the U-70 accelerator at IHEP”. The measurements were supported by the set of extensive simulations using the CERN FLUKA code. Good agreement between the FLUKA simulations and measurements was shown. Even single monitor with lead insertion calibrated at AmBe neutron source allows to have a good result in the real neutron field. The possibility to estimate the fluence of the high energy neutrons with the energy above 10 MeV using data from the pair of monitors was demonstrated. To improve the quality of the measurements one needs to take into account difference between calibration and measurements conditions.

About the Authors

I. L. Azhgirey
Institute for High Energy Physics named after A. A. Logunov of National Research Center “Kurchatov Institute”
Russian Federation

Igor L. Azhgirey, Candidate of Physical and Mathematical Sciences, Head of the Laboratory

Protvino



I. S. Bayshev
Institute for High Energy Physics named after A. A. Logunov of National Research Center “Kurchatov Institute”
Russian Federation

Igor S. Bayshev, Candidate of Physical and Mathematical Sciences, Leading Researcher 

Protvino



V. A. Pikalov
Institute for High Energy Physics named after A. A. Logunov of National Research Center “Kurchatov Institute”
Russian Federation

Vladimir A. Pikalov, Head of the Laboratory

Protvino



O. V. Sumaneev
Institute for High Energy Physics named after A. A. Logunov of National Research Center “Kurchatov Institute”
Russian Federation

Oleg V. Sumaneev, Researcher

Protvino



References

1. Dabrowski A., Stickland D., Auzinger G., Pasztor G., Lujan P. et al. (CMS Collaboration). The Phase-2 Upgrade of the CMS Beam Radiation Instrumentation and Luminosity detectors. CERN, Geneva, CERN-LHCC-2021-008, CMS-TDR-023, 2021.

2. Azhgirey I. L., Bayshev I. S., Kurochkin I. A., Pikalov V. A., Sumaneev O. V., Lukanin V. S. Neutron monitors for high energy accelerators. In: Proceedings of 26th Russian Particle Acc. Conf. RuPAC2018, Ptovino, Russia, 2018, p. 225, DOI:10.18429/JACoW-RUPAC2018-TUPSA38

3. Gribushin A. M. et al. A Neutron Field Monitoring System for Collider Experiments. Instruments and Experimental Techniques, 2017, Vol. 60, No. 2, pp. 167–174. DOI: 10.1134/ S0020441217020051

4. The CMS Collaboration. The Phase-2 Upgrade of the CMS Beam Radiation, Instrumentation, and Luminosity Detectors: Conceptual Design. CERN, Geneva, CMS NOTE-2019/008

5. Sumaneev O. V., Azhgirey I. L., Baishev I. S., Pikalov V. A. Neutron field Measurements by GFPC based monitors at the Carbon beam at IHEP U-70 proton synchrotron. In: Proceedings of 27th Russian Particle Acc. Conf. RuPAC2021, Alushta, Russia, p. 129, 2021. DOI:10.18429/JACoW-RuPAC2021-FRC01

6. Antipov Yu. M., Britvich G. I., Ivanov S. V. et al. Formation of a transversely flat dose field and the first radiobiological experiments on a carbon beam extracted from the U-70. Instruments and experimental techniques, 2015, No. 4, pp. 107–116. (in Russ.)

7. Ahdida C. et al. New Capabilities of the FLUKA Multi-Purpose Code. Frontiers in Physics, 2022, No. 9, p. 788253.

8. Battistoni G. et al. Overview of the FLUKA code. Annals of Nuclear Energy, 2015, No. 82, pp. 1018.

9. INTERNATIONAL STANDARD ISO 8529-1. Neutron reference radiations fields – Part 1: Characteristics and methods of production (ISO 8529-1:2021, IDT).